In stock
Warranty: 7 to 10 Business Days
Condition: Refurbished and Tested to Manufacturer Spec
Agilent 1100 HPLC System with Variable Wavelength Detector & Fluorescence Detector
The Agilent 1100 HPLC System guarantees the most efficient mixing and pulse-free solvent delivery for consistently fast and accurate results. The system’s optimized design using a quaternary pump allows variable stroke volumes for stable solvent flow. This electronically activated inlet valve runs synchronous to the pistons to eliminate vapor formation in volatile solvent, making the Agilent 1100 both remarkably easy-to-use and maintenance-free. It also features a high-speed proportioning valve and a pre-configured degasser, which allow convenient low-pressure mixing and a large solvent choice flexibility. The HPLC System’s vacuum degasser ensures reduced operating costs without the use of helium and optimum instrument performance.
The Agilent 1100 Variable Wavelength Detector lowers detection limits with simplified optics for high sensitivity, low baseline drift, and wide linear range. This wide linear range allows for reliable simultaneous quantitation of main and side products and impurities. The Agilent 1100 VWD’s minimum baseline noise gives it the highest baseline stability for the lowest detection limits. This VWD offers programmable wavelength switching for sensitivity and optimized selectivity optimized to each analyte’s elution, along with stop-flow wavelength scanning for quick wavelength optimization.
The Agilent 1100 Fluorescence Detector brings high-sensitivity and selection through simultaneous multi-wavelength detection. This easy-to-use detector provides quantitative data and fluorescence spectra from a single run. The FD's spectral information can be used for rapid method optimization and verification of separation quality.
The unit can be operated with a hand-held controller or computer running Clarity software. The HPLC System features extensive diagnostics with compatible data analysis software for on-site and remote trouble-shooting.
All of our equipment is fully tested to perform at factory specifications before being shipped.
Refurbished Agilent 1100 System Includes:
G1311A Quat Pump (Binary Pump Available)
G1313A Autosampler (G1329A Thermal Autosampler Available)
G1314A VWD Detector
G1321A FLD Detector
G1316A Column Compartment
G1322A Degasser (G1379a Microdegasser Available)
G1323B Gameboy Controller
Windows 7 Data Station with Monitor
Chemstation version 4.3 software
All Power Cords > Can/Crossover Cables > Lan Interface Card > Solvent Tubing > Capillaries > Full 90 Day Parts and Labor Warranty
Agilent 1100 HPLC System Specifications
Type
|
Specification
|
Detection type
|
Double-beam photometer
|
Light Source
|
Deuterium lamp
|
Wavelength Range
|
190 - 600 nm
|
Short term noise (ASTM) Single
and Multi-Wavelength
|
± 0.75 × 10-5AU at 254 nm
|
Drift
|
3 × 10-4AU/hr at 254 nm
|
Linear Absorbance range
|
> 2 AU (5%) upper limit
|
Wavelength accuracy
|
± 1 nm
|
Slit width
|
1, 2, 4 , 8, 16 nm
|
Band Width
|
6.5 nm typical
|
Diode Width
|
< 1 nm
|
Flow Cells
|
Standard: 14-µl volume, 10-mm
cell path length and 40 bar
(588 psi) pressure maximum
High pressure: 14-µl volume,
10-mm cell path length and
400 bar (5880 psi) pressure
maximum
Micro: 1-µl volume, 5-mm cell
path length and 40 bar (588 psi)
pressure maximum
Semimicro: 5-µl volume, 6-mm
cell path length and 40 bar
(588 psi) pressure maximum
|
Control and Data Evaluation
|
Agilent ChemStation for LC
|
Analog Outputs
|
Recorder/integrator: 100 mV or 1 V,
output range 0.001 – 2 AU, two
outputs
|
Communications
|
Controller-area network (CAN), GPIB,
RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN
optional
|
Safety and Maintenance
|
Extensive diagnostics, error detection
and display (through control module
and ChemStation), leak detection, safe
leak handling, leak output signal for
shutdown of pumping system. Low
voltages in major maintenance areas.
|
GLP Features
|
Early maintenance feedback (EMF) for
continuous tracking of instrument
usage in terms of lamp burn time with
user-settable limits and feedback
messages. Electronic records of
maintenance and errors. Verification of
wavelength accuracy with built-in
holmium oxide filter.
|
Housing
|
All Materials recyclable
|
Agilent 1100 Series HPLC System Operator's Manual