Features: Fluorescence, Luminescence, Visible Absorbance, UV absorbance, Stacker, WorkOut2 Data Analysis Software.
General Features:
Light Sources:
- Tungsten-halogen lamp
- For fluorescence and absorbance measurements.
- Xenon flash lamp
- For TR-fluorescence and UV absorbance measurements.
Detection units:
- Photomultiplier tube
- For fluorescence, luminescence and TR-fluorescence measurements.
- Photodiode
- For absorbance measurements.
Plate Formats: 1 to 1536
Shaking: Linear, orbital and double orbital. Duration, speed and amplitude are adjustable.
Temperature Control:
- Temperature range: from 2 °C above ambient up to 45 °C (0.1 °C increments)
- Uniformity: ± 1 °C
- Heating time: < 10 minutes (from RT up to 37 °C)
Dispenser: The dispenser module consists of 1-4 high quality syringes. Depending on the dispenser option, up to 4 separate channels can be directed to one well (96-well plate).
Dispense volume: 5-350 µL (1 µL increments)
Accuracy typically: < 5 % for 5 µL, < 0.5 % for 50 µL, < 0.05 % for 350 µL
Precision typically: < 1.4 % for 5 µL, < 0.2 % for 50 µL, < 0.02 % for 350 µL
Speed: Minimum and maximum speeds are related to the volume.
Dead volume: < 0.5 mL
Specifications:
Fluorometry: 96-well black plate (200 µL)
- Fluorescein: typically < 2 fmol/well, 10 pM
- Linearity: > 5 decades
- Crosstalk: < 0.01 %
- Umbelliferone: typically < 200 fmol/well, 1 nM
- Rhodamine: typically < 100 fmol/well, 0.5 nM
TR-Fluorometry: 96-well clear plate (200 µL)
- Europium: typically < 10 amol/well, 50 fM
- Linearity: > 5 decades
- Crosstalk: < 0.01 %
- Terbium: typically < 5 amol/well, 25 fM
- Samarium: typically < 50 amol/well, 250 fM
- Dysprosium: typically < 150 amol/well, 750 fM
Fluorescence polarization
- Fluorescein: 1 nM, 40 µL
- 384-well black plate Standard deviation: < 5 mP
Luminometry: 96-well plate
- Lower limit of detection (LLD): total flux of 20,000 photons/s (standard PMT)
- 100,000 photons/s (red sensitive PMT)
- typically 1 amol/well with
- ALP: AMPPD substrate
- ATP: typically 80 amol/well in flash assays
- Luciferase: 0.9 pg/well using steadylite
Photometry: 96-well plate
- Measuring range @ 405 nm: 0-4 A
- Accuracy @ 405 nm: < 2 % (or 0.01 A) within 0-2 A
- Precision @ 405 nm: < 0.5 % (or 0.01 A) within 0-2 A
Dimensions
- Height: 383 mm (15.5"); with stackers 510 mm (20")
- Depth: 590 mm (23.5")
- Width: 485 mm (19")
- Weight: 49 kg (108 lbs); with stackers 59.5 kg (131.2 lbs)
Power Requirements
- Power consumption: 250 VA
- Main voltage: 110-120 V/220-240 V, 50/60 Hz