Ambulatory Infusion Pump
Product Specifications:
Pump Resolution - Standard reservoir or set = 0.050ml/pump stroke nominal. High volume set = 0.100 ml/pump stroke nominal
Dimensions - 1.7" D x 4.1" W x 5.6" H, excluding cassette
Weight - 20 oz including 9 volt battery and empty 100 ml reservoir
Power Source - 9 volt alkaline or lithium battery, External Power Source, or AC Adapter
System Delivery Accuracy - ±6% nominal
PCA Delivery Mode Specifications
Reservoir Volume - 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
Units - ml, mg, mcg
Concentration - mg/ml, mcg/ml
Continuous Rate - 0-30 ml/hr or the mg or mcg equivalent
Demand Dose - 0-9.9 ml, Delivery rate is 125 ml/hr nominal
Demand Dose Lockout:
5 min-24 hours in the following increments:
- 1 min for values between 5 and 20 minutes
Max Doses Per Hour - 1-12 doses in 1 dose increments
Demand Doses Given - 0-999
Demand Dose Attempts - 0-999
Given - 0-99999.99 in 0.01 unit increments
Clinician Bolus - 0.1 ml - 20.0 ml or mg or mcg equivalent, delivery rate is 125 ml/hr nominal
Continuous Delivery Mode Specifications
Reservoir Volume - 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
Continuous Rate:
0.1-350 ml/hr in the following increments:
- 0.1 for values between 0.1-100
-1 for values between 100-350
Given - 0-99999.9 in 0.1 unit increments
TPN Delivery Mode Specifications
Reservoir Volume - 10.0 - 9990 or Not in Use; programmable in 1 ml increments, displayed in 0.1 increments
Infusion Volume - 10-9990 ml in 10 ml increments
Infusion Period - 0 hrs 10 min - 99 hrs 40 min in 10 minute increments
Taper Down Period - 0 hrs 0 min - 99 hrs 50 min in 10 minute increments
Plateau Rate - Calculated by pump; 10-350 ml/hr
KVO Rate - Calculated by pump; 1/10 of Plateau Rate up to 5 ml/hr
Given - 0-99999 in 0.1 unit increments
Intermittent Delivery Mode Specifications
Reservoir Volume - 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
Dose Volume:
0.1-1000 ml in the following increments:
0.1 for values between 0.0 and 100
1 for values between 100 and 1000
Dose Duration:
1 min-24 hrs in the following increments:
1 min for values between 1 min and 10 min
5 min for values above 10 min
*Duration is limited by Dose Volume so that rate does not exceed 350 ml/hr
Dose Cycle - 10 min - 96 hrs in 5 minute increments
KVO Rate: - 0-10 ml/hr in 0.1 ml increments
Next Dose Start Time - 10 min - 96 hrs X min (where X equals a 10 minute increment in the 96th hour)
Given - 0-99999.9 in 0.1 unit increments
Time Remaining - Dose and Cycle display in 1 min increments