Range Calibrator w/option:
H01= Ext. Programmable
DC Reference Capability
The Agilent (now Keysight) 11683A range calibrator is specifically designed for use with the EPM, EPM-P series, E1416A and the discontinued 70100A and 43X power meters.
It allows verification of full-scale meter readings on all ranges, as well as meter tracking. Simply connect the cable between the power meter and calibrator.
The CAL ADJ control on the power meter is used to set the meter to full scale on the
1 mW range.
The 11683A with option H01 is a standard 11683A range calibrator that has been modified so that it will operate with eihter the internal DC reference source or an external programmable DC reference source.
Verification of full-scale meter readings on all ranges and meter tracking
Polarity switch that tests the auto-zero circuit
Calibration functions:
Outputs corresponding to meter readings of 3, 10, 30, 100, and 300 µW; 1, 3, 10, 30, and 100 mW
Calibration Uncertainty:
+/- 0.25% in all ranges
Power: 100, 120, 220, or 240 Vac +5% -10%, 48 - 440 Hz, <10 VA
Size (WxHxD): 133 x 89 x 216 mm (5.25 x 3.5 x 8.5 in.)
€ 2,190 (net) $2,759 (US)