Tektronix AM503 Current Probe Amplifier

PRODUCT SPECS

(From Used-Line T&M Specifications )
Status Discontinued

The AM503 current-probe amplifier allows display of current on any oscilloscope having 10 mV/div sensitivity, 50 Ω or 1 MΩ input, and (for performance to full bandwidth) at least 100 MHz bandwidth. The amplifier attenuator has 12 calibrated steps in a 1-2-5 sequence, and the knob-skirt is illuminated to indicate current / division. The AM503 forms current probe systems with the A6302 and A6303.

The AM503S package includes the AM503 with the A6302 or the A6303 or both.

  • Maximum input current (dc + peak ac): 20 A with A6302, 100 A with A6303
  • Maximum voltage for current under test - bare conductor (dc + peak ac): 500 V with A6302, 700 V with A6303
  • -3 dB bandwidth: dc to at least 50 MHz with A6302, dc to at least 15 MHz with A6303
  • Full bandwidth rise time: 7 ns or less with A6302, 23 ns or less with A6303
  • Deflection factor: 1 mA/div to 5 A/div for A6302, 20 mA/div to 50 A/div for A6303

CONTACT THE SELLER (RFQ)


To:  Liberty Test Equipment, Inc.
 
All fields are required
 Remember me
 Send me a copy.
Send
News, sales, & auction announcements
About the Seller
since 2006



Liberty Test Equipment, Inc.


815
Visitors viewed Liberty Test Equipment, Inc. ads in the last 30 Days.


The dealer's latest update: over 30 days


1640 Lead Hill Blvd. Suite 120 Roseville 95661 CA USA  
More about this seller »


Sales Conditions:

END USERS ONLY! 10-day right of return, 12-month warranty, around-the-clock NIST traceable/ISO registered lab & all standard accessories. Pricing and availability are subject to change. Many payment options, terms, credit card, checks, lease terms, rental, etc.

Some of the services we offer:

(Rentals) (Calibration) (Repair) (Manuals) (Market place)

Similar Listings: